continuous burn- and carriergassupply for TOC-analyzers
enables exact analysis in case of flat basiclines
complete selfcontrolling system by intelligent electronic
save operation by low working pressure
TOC-gas generation
6 new models of the series TG are offered to produce UHP-air for TOC-analyzers. All models are working nearly wear-resistant and can be installed in an easy way. The generators are producing continuous burn- and carriergases with following concentrations: HC < 0,1 ppm (measured as methane), CO2 < 1 ppm and H2O < 1 ppm. Customer`s air (6 bars recommended) will be prepared in the generator with modern filtertechnology to eliminate microparticle, water and oilmist. The following totally catalytic oxidation separates hydrocarbons including the methane. Now the CO2- and watersteamtraces will be eliminated by the pressure swing adsorption system, so the requirements of TOC-analyzer manufactorer will be performed. The maximum flowrate, the gas-/airsupply and the nominal temperature in the reactor can optionally be controlled as well as the powersupply can be supervised. In case of stoppage or an exceed of the limit an alarmsignal will be reported over a relay, corresponding to ISO 9000. Is a defect recognized the generators do a switch over to gasbottles or they lock the supply to the laboratory. For maintenance the prefilter should be changed yearly. 10 minutes after switch-on the biggest generator produces up to 20 l/min burn- and carriergas with constant quality. Maximum inlet pressure 6-10 bars for optimum workingpoint. Pressure differences of < 1 bar by the maiximum flowrates. Please ask seperately technical data for systems with integrated compressors.
Compressed air supply
Doesn´t exist any compressed air supply all models are available with external low noise (< 55 dbA) compressors.
Benefits
Strictly flat basiclines without any basicnoise permit high performed results in traceanalysis. The gasqualitiy is comparable with UHP-qualitiy gases, but extremely cheaper. Low working pressures (max. 10 bars) permit the installation of the gernators in laboratories, directly by analyzers. Difficult transports of hevy gasbottles are not necessary anymore. The independence of gassupplier makes more place in laboratories possible and saves working-time. The advanced serie TG-SMART recognizes not only defects in the periphery and reports them but also it reacts correspondingly and switches by stoppage of voltage for example automatically over to a reserve bottle.
Technical Data
Typ
TG 600
TG 2000
TG 4000
TG 8000
TG 15 K
TG 20 K
Max. gasflow(ml/min)
600
2000
4000
8000
15000
20000
CO2- and H2O-concentration at the gas outlet
< 1 ppm CO 2, < 1 ppm H2O
Max. HC-concentration at the gasoutlet
< 0,1 ppm (as CH4)
Gasconnections (swagelok)
6mm
Power supply (Wh)
230 V/50 Hz
Starting time (min)
Approx. 10
Power demand
< 30
< 40
< 60
< 90
< 120
< 150
Max. power input (W)
420
1600
Dimensions (W x H x D)
310 x 540 x 460mm
470 x 540 x 460mm
Shipping weight (kg)
10
12
15
20
24
30
Options
flowmeter
pressure display/manometer
temperature OK / ready (LED-display reactortemperature)
isolated alarmcontacts without logfunction stoppage of power supply stoppage of gas-/air supply (1-8 bars adjustable) nominal temperature remain under max. gasflowrate overflow
electrally isolated alarmcontacts with logfunction Malfunctions will be located by electronic (by example "malfunction air supply"). The malfunctions will be displayed and signalized to the connectors by relay. After the malfunction is corrected the error has to be confirmed manually and the generator begins to function.
consequences of stoppage gasflow will be closed up switch over to gasbottles