The instrument meets the generic qualityconsistency evaluation
RCY-1400T (Test status)
Registered trademark: Haiyida
Proprietary product
In summarizes many year productions tothe dissolution tester in success empirical base, RCY-1400T dissolution testeris our company uses advanced technology and components to decelop a new drugdissolution tester. The instrument adopts the high quality LCD touch screen, 14cups 12 poles with double row. The "main menu" and "basicexperiment" option can be convenient to the conventional repetitive trialsor research.
Features:
? Using highquality LCD touch screen, the operation interface intuitive and easy to be used
? Produced byelectric lift
? On center polewith dissolution cup automatic positioning
? Standardconfigurations by fourteen 1000ml dissolution cups, 12 basket poles and 12paddles, and have 2 rehydration insulation cup special holes
? Temperature sensorwith high precision temperature control, adopts dual temperature sensor,temperature control, the other temperature measuring all the way, could detectthe dissolution temperature at any time
? With functions oftemperature control self-inspection
? Basket, paddle andsmall cup method from the bottom of the cup height automatic positioning
? Has the usermanagement function, logins user account by the user name and password, canundertake various parameters under this account set up and run the test
? Storage can bepreset 27 groups of experimental parameters
? Can be setequipment calibration date, expire automatically prompt
? Timing shutdown,timing, automatic preheating function
? Date, real-timeclock, temperature calibration function
? Ultra, ultra hightemperature alarm and secondary overheating protection function
? Can store the userlogin information and the test record ? Optional miniature printer real-time test data and historical records
Specifications:
? Wobble of paddle: ≤0.5mm
? Wobble of rotatingbasket: ≤1.0mm
? Deviation betweenthe shaft and the vessel: ≤2.0mm
? Speed control:Adjustable range: 10 to 300 rpm
? Speed resolution:1rpm
? Speed accuracy: ≤±4%
? Temperaturecontrol: Adjustable range: 5℃(room temperature) to 55℃
? Temperatureresolution: 0.01℃
? Temperatureaccuracy: ≤±0.3℃
? Methods storage:27 groups
? Number of samplingintervals: maximum 25
? Sampling cycletime: maximum 99 hours 59 minutes per cycle
? Working noise: <60dB
? Power requirement:AC 220V±10% 50 Hz, 1350W
? Environmentrequirement: temperature 5 to 37℃, humidity ≤80%
? Safetyperformance: in line with national industry standards